|
All inorganic scintillators are non-proportional. This means that
the amount of scintillation light generated in response to
energy deposited by a γ-ray is not strictly
proportional to the deposited energy. Aside from
photo-electron statistics, and especially at energy deposits
beyond 1 MeV, this is a limiting factor for the energy
resolution any scintillator can achieve.
One way to quantify this phenomenon is to compute the
apparent brightness of the scintillator as a function of
deposited energy. Here we measure brightness in units of PMT
anode pulse charge vs deposited energy. We arbitrarily define
the value at 662 keV as 100% and quantify the
non-proportionality by how much the charge-to-energy ratio deviates
from the pivot point at 662 keV.
Figure
3: Non-proportionality as measured in a 3-inch CsI(Na)
crystal. The general trend is that the CsI(Na) crystal
appears brighter at lower γ-ray energies and loses
luster as the γ-ray energy increases. Note that this
non-proportionality is a function of the γ-ray energy, not
the total deposited energy. Cascade sums are not subject to
scintillator non-proportionality.
60Co is the perfect example for this phenomenon.
Place a 1 μCi (37 kBq) 60Co-source
right against the front face of a 3-inch radiation sensor.
The excited daughter of 60Co emits an
1172 keV and a 1333 keV γ-ray in coincidence.
When both are fully absorbed in the scintillator, the sum
signal should precisely match the formula 1172 keV +
1333 keV = 2505 keV.
Note that the non-proportionality is mostly, but not exclusively a
characteristic of the material. Especially at the low energy
end the effect is more pronounced in small crystals due to
geometry and light collection effects.
|